Z-1139-2024: BIOFIRE Blood Culture Identification 2 (BCID2) Panel when used with: BD BACTEC Lytic Anaerobic medium BD

Recall record Z-1139-2024 for BioFire Diagnostics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1139-2024
Company in FDA record
BioFire Diagnostics, LLC
FDA classification date
Feb 15, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a trend in false positive Candida tropicalis results on blood culture panel when using certain BD BACTEC vial mediums

Source facts

Action
On 01/26/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via overnight FedEx to customers to inform them of the increased risk of false positive Candida tropicalis results when the BIOFIRE BCID2 Panel is used with BD BACTEC blood culture vials including, but not limited to, the bottle types: BD BACTEC" Lytic Anaerobic medium BD BACTEC" Peds Plus medium BD BACTEC" Plus Aerobic medium BD BACTEC" Plus Anaerobic medium BD BACTEC" Standard Aerobic medium BD BACTEC" Standard Anaerobic medium Customers are requested to take the following actions: - If the BIOFIRE BCID2 Panel is used to test BD BACTEC blood culture vials, positive results for Candida tropicalis should be confirmed by another method prior to reporting the test results. - Distribute this information to all appropriate personnel in their laboratory, retain a copy in their files, and forward this information to all parties that may use this product, including others to whom they may have transferred the product. - Complete the Acknowledgement Form in Attachment A and return it to bioMerieux so that bioMerieux may acknowledge their receipt of this notification. It is important that they return the acknowledgement form to bioMerieux even if they determine that this urgent field safety notice does not impact their facility. For questions/assistance contact BIOFIRE Technical Support Team at biofiresupport@biomerieux.com or via telephone at 1-800-735-6544 and select option 5.
Initiated
2024-01-26
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Andorra, United Arab Emirates, Armenia, Argentina, Austria, Australia, Bosnia, Belgium, Bulgaria, Bahrain, Brunei, Bolivia, Brazil, Canada, Switzerland, Cote D'Ivoire, Chile, China, Colombia, Costa Rica, Cyprus, Cech Republic, Germany, Denmark, Dominican Republic, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Gabon, United Kingdom, Georgia, Guadeloupe, Greece, Guatemala, Hong Kong, Honduras, Croatia, Hungary, Iceland, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, Korea, Kuwait, Lebanon, Lithuania, Latvia, Morocco, Macedonia, Mali, Mongolia, Macau, Mexico, Malaysia, New Caledonia, Nigeria, Nicaragua, Netherlands, Norway, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, St. Pierre/Miquelon, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, El Salvador, Thailand, Tunisia, Turkey, Taiwan, Uganda, Uruguay, Vatican City, Venezuela, Vietnam, South Africa, and Zimbabwe.
Recall number
Z-1139-2024
Recall status
Open, Classified
CDRH record ID
204997
FDA classified
2024-02-15
Recall event number
93636

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗