Open original FDA record ↗Recall record
- FDA identifier
- Z-1140-2022
- Company in FDA record
- MEDLINE INDUSTRIES, LP - Northfield
- FDA record creation date
- May 26, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this impacted Calcium reagent will not be provided due to probable Quality Control failure. A delay in diagnosis may be experienced with repeat testing resulting in patient (customer) inconvenience, pain, and cost of the repeat blood draw and testing. All HII1220001954 product shipped/stored at a Medline branch between July 2021 and August 2021 in the identified lots were incorrectly stored and shipped. Six (6) boxes have been shipped to end users at room temperature instead of refrigerated (ice). Five (5) boxes are on hold at a Medline warehouse.
Source facts
- Action
- Medline contacted affected consignee via telephone on 10/22/2021 and confirmed that the product had been destroyed.
- Initiated
- 2021-10-22
- Root cause
- Process control
- Terminated
- 2023-01-04
- Distribution
- US Distribution to North Carolina.
- Recall number
- Z-1140-2022
- Recall status
- Terminated
- CDRH record ID
- 193532
- FDA record created
- 2022-05-26
- Recall event number
- 90126