HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no
Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All
other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the
validated linearity range for each assay.
Source facts
Action
A recall letter, Field Safety Notice & Acknowledgement mailed to the customers March 20, 2013. The letter included the product information, reason for recall, instructions on what to do with the recalled product, and contact information:"If you have any questions or concerns, please contact your local HORIBA Medical
representative. We sincerely apologize for any inconvenience this may cause your laboratory. Please contact your Customer Service Representative at 1-888-903-5001 (option 3) for any additional questions related to this customer
letter."
Initiated
2013-03-20
Root cause
Software design
Terminated
2014-09-26
Distribution
Nationwide Distribution.
Recall number
Z-1141-2013
Recall status
Terminated
CDRH record ID
116862
FDA classified
2013-04-18
Recall event number
64723