Open original FDA record ↗Recall record
- FDA identifier
- Z-1141-2021
- Company in FDA record
- Ecolab Inc
- FDA record creation date
- Feb 22, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Slits were identified in a low quantity of ORS-320 Slush + Warmer Disc Drapes from a single lot. A source for potential manufacturing error was identified and investigation is underway to prevent further recurrence. The slits may lead to a breach in the sterile barrier between fluid in the basin and the slush/warming equipment. This could pose a risk of infection to patients receiving warmed irrigation fluid or surgical slush if used during a procedure.
Source facts
- Action
- The firm sent a recall communication letter "Urgent: MEDICAL DEVICE RECALL IMMEDIATE ATTENTION REQUIRED" to customers on 01/11/2021.
- Initiated
- 2021-01-11
- Root cause
- Process control
- Terminated
- 2023-01-09
- Distribution
- Worldwide distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, FL, HI, ID, IL, IN, KS, LA, MA, MD, MI, MO, MT, NJ, NY, OH, OK, OR, PA, TN, TX, UT, WA. In the country of Japan.
- Recall number
- Z-1141-2021
- Recall status
- Terminated
- CDRH record ID
- 185485
- FDA record created
- 2021-02-22
- Recall event number
- 87173