Z-1141-2023: Life2000 Ventilator PKG, Hospital, a) REF BT200011, b) REF BT-20-0011, ventilatory support for the care o

Recall record Z-1141-2023 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1141-2023
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Feb 24, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.

Source facts

Action
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its healthcare providers, wholesalers, distributor/reseller, and durable medical equipment providers on 01/25/2023 by USPS First Class Mail. The notice explained the issue and the hazard and requested the following actions be taken by adressee: "Please follow the daily checks and preventive maintenance requirements below to ensure the best oxygen delivery with the Life2000 ventilation system when used with a third-party oxygen concentrator. 1- Daily Checks a) Check the CombO2 hose and Breathe Pillow Interface for kinks, dirt, moisture, or damage and contact Baxter Customer Service at 800-426-4224, option 3, if replacement parts are needed. Baxter recommends replacing the Breathe Pillow Interface every 90 days and the CombO2 hose every six months. b) Do not modify, alter, or extend tubing, connectors, or interface. Additionally, when using the Life2000 system with an oxygen concentrator, only use Baxter s approved 6-foot, 20-foot or 50-foot CombO2 hose. c) Address alarms as described in the Life2000 Quick Reference Guide (PL-20-0044-D) pages 17-19. d) Make sure the CombO2hose is correctly attached to the Life2000 compressor and the oxygen concentrator, as shown in Attachment A, Figure 1. Make sure all connections are tight. e) Ensure the setting on the oxygen concentrator remains at the prescribed liter flow while connected to the Life2000 system as shown in the Life2000 Extended Range Configuration Quick Start Guide (APR186301). If the liter flow drops when you connect to the Life2000 system, take the following steps: - Stop using the Life2000 system and go back to your traditional oxygen cannula and concentrator. - Call Baxter s Clinical Support team at 800-397-9071 for troubleshooting. f) Please notify your health care team of any concerns or seek medical attention if you are exhibiting signs of respiratory distress (increased breathing rate, wheezing, bluish color around the mouth, lips or fingernails,
Initiated
2023-01-25
Root cause
Under Investigation by firm
Distribution
Nationwide
Recall number
Z-1141-2023
Recall status
Open, Classified
CDRH record ID
198080
FDA classified
2023-02-24
Recall event number
91506

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗