Open original FDA record ↗Recall record
- FDA identifier
- Z-1142-06
- Company in FDA record
- Disetronic Medical Systems, Inc.
- FDA classification date
- Jul 8, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
Source facts
- Action
- "U.S. and International distributors, patients and their physicians were notified via letters dated 3/31/06. Distributors were instructed to search for all part numbers affected by the recall and to provide all affected customers and sub-distributors with the recall notice. Customers were offered a replacement set or replacement with a different model. A press release was also issued on 4/3/06.
- Initiated
- 2006-03-31
- Root cause
- Other
- Terminated
- 2007-11-06
- Distribution
- Worldwide including USA, Argentina, Austria, Barbados, Belgium, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Honduras, Hungary, Italy, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
- Recall number
- Z-1142-06
- Recall status
- Terminated
- CDRH record ID
- 45917
- FDA classified
- 2006-07-08
- Recall event number
- 34905