Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.
Source facts
Action
On June 29. 2018, the firm sent Customer Information Letters to affected customers. The letter informed customers that Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images.
Customers were advised to review the user documentation provided with your system related to Quality Assurance checks. If artifacts are detected during review of the images produced by these checks, and are not removed with the recommended steps identified within your user documentation, please contact your service representative before resuming clinical use.
Initiated
2018-07-03
Root cause
Device Design
Terminated
2020-04-17
Distribution
Distributed only to China and UK.
Recall number
Z-1142-2019
Recall status
Terminated
CDRH record ID
165989
FDA record created
2019-04-13
Recall event number