Open original FDA record ↗Recall record
- FDA identifier
- Z-1143-2021
- Company in FDA record
- Natus Manufacturing Limited
- FDA record creation date
- Feb 22, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The external drainage system failed to meet Pyrogen test requirements for devices contacting cerebrospinal fluid of 2.15 endotoxins.
Source facts
- Action
- On January 12, 2021, Natus sent an "Urgent Field Safety Notice - Recall" notification to all affected consignees via US Mail. In addition to informing consignees about the recall, the notification asked consignees to take the following actions: 1. Please return these affected items at your earliest convenience to the following address: 3150 Pleasant View Road, Middleton, WI 53562 USA 2. Replacement product is available. The Technical Service department will be in contact in relation to the provision of the replacement product. 3. Please complete and return the customer reply form to Natus at the following Email address: FCA@Natus.com
- Initiated
- 2021-01-12
- Root cause
- Release of Material/Component prior to receiving test results
- Terminated
- 2022-11-08
- Distribution
- US Nationwide distribution in the state of NY.
- Recall number
- Z-1143-2021
- Recall status
- Terminated
- CDRH record ID
- 185486
- FDA record created
- 2021-02-22
- Recall event number
- 87175