Skip to content

Public FDA record

Z-1143-2021: Ref: NT821731C, Natus EDS 3 CSF External Drainage System - Product Usage: indicated for draining cerebro

Recall record Z-1143-2021 for Natus Manufacturing Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1143-2021
Company in FDA record
Natus Manufacturing Limited
FDA record creation date
Feb 22, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The external drainage system failed to meet Pyrogen test requirements for devices contacting cerebrospinal fluid of 2.15 endotoxins.

Source facts

Action
On January 12, 2021, Natus sent an "Urgent Field Safety Notice - Recall" notification to all affected consignees via US Mail. In addition to informing consignees about the recall, the notification asked consignees to take the following actions: 1. Please return these affected items at your earliest convenience to the following address: 3150 Pleasant View Road, Middleton, WI 53562 USA 2. Replacement product is available. The Technical Service department will be in contact in relation to the provision of the replacement product. 3. Please complete and return the customer reply form to Natus at the following Email address: FCA@Natus.com
Initiated
2021-01-12
Root cause
Release of Material/Component prior to receiving test results
Terminated
2022-11-08
Distribution
US Nationwide distribution in the state of NY.
Recall number
Z-1143-2021
Recall status
Terminated
CDRH record ID
185486
FDA record created
2021-02-22
Recall event number
87175

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗