Skip to content

Public FDA record

Z-1143-2024: Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement

Recall record Z-1143-2024 for Voco GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1143-2024
Company in FDA record
Voco GmbH
FDA classification date
Feb 16, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.

Source facts

Action
Voco issued a field safety notice to its consignees on 01/25/2024 via email. The notice explained the problem, risk, and requested the product be returned. Those who have sold the product were directed to notify their customers and collect the affected product. For questions contact Mr. Michael Cyll at m.cyll@voco.de.
Initiated
2024-01-25
Root cause
Process control
Distribution
Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland.
Recall number
Z-1143-2024
Recall status
Open, Classified
CDRH record ID
205804
FDA classified
2024-02-16
Recall event number
93877

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗