- FDA identifier
- Z-1144-2021
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Feb 23, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Misleading error messages and a gap in the Operator Manual which affects ARTIS Icono biplane or ARTIS Icono floor systems with software version VE20B. Potential issues include System error management, Erroneous error messages, Zoom/Pan Function, Grid Indication, and Coolant Level. May cause procedures to be terminated and performed on an alternative x-ray system
Source facts
- Action
- Siemens Medical Solutions USA, Inc. issued a Customer Safety Advisory Notice to all affected customers via AX080/20/S. on 12/28/20. Letter states reason for recall, health risk and action to take: Siemens will provide a software update to all affected customers which will correct the issue via Update Instruction AX070/20/S. The Operator Manual addendum to the Instruction for Use for VE20 has been updated and includes the missing information about the grid indication Our service organization will contact you to arrange a date to perform this corrective action. Please feel free to contact our service organization at 1-800-888-7436.
- Initiated
- 2020-12-18
- Root cause
- Software design
- Terminated
- 2021-11-17
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1144-2021
- Recall status
- Terminated
- CDRH record ID
- 185480
- FDA record created
- 2021-02-23
- Recall event number
- 87169