Z-1145-2019: Bilirubin Total Gen.3, Cat. Nos. 05795397190, 05795419190, 05795320160, 05795338160, 05795346160 ***adde

Recall record Z-1145-2019 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1145-2019
Company in FDA record
Roche Diagnostics Corporation
FDA classification date
Apr 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***

Source facts

Action
On 12/17/18, the firm distributed an Urgent Medical Device Correction letter to affected consignees. Customers were notified of the interference issues. The Limitations-interference section of the method sheets for the cobas c 311 analyzer and the cobas c 501, 502, 701, and 702 modules are being updated with the following claims: Bilirubin Total Gen.3 assay: Immunoglobulins: No significant interference from immunoglobulins up to a concentration of 28 g/L (187 mol/L) (simulated by human immunogloblulin G). Lactate Gen.2 assay (CSF samples): Ditaurobilirubin: No significant interference from ditaurobilirubin up to an approximate concentration of 102 mol/L (6 mg/dL). Tina-quant IGM-2 (Sensitive application): Hemolysis: No significant interference up to an H index of 600 (approximate hemoglobin concentration of 373 mol/L or 600 mg/dL). Lipemia (Intralipid): No significant interference up to an L index of 600 Customers were asked to do the following: " Be aware of the new interference claims listed in this Urgent Medical Device Correction (UMDC). This UMDC serves as labeling until the updated method sheets are available. The method sheets for the MODULAR ANALYTICS P/D modules will not be updated. " Manually update the H and L indexes for the IGMP2 sensitive application on your cobas c 311 analyzer or cobas c 501, 502, 701, and 702 modules following the instructions provided in this UMDC. " Complete all sections of the enclosed fax form (TP-00507) and fax it to 1-xxx-xxx-xxx or email it to rochexxxx@stericycle.com. " File this UMDC for future reference. When the method sheets are updated, they will be available on the diagnostics.roche.com Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions. website and on the cobas e-library. ***UPDATE: Firm sent updated communication to customers dated 5/9/2019 informing them that L and H index values
Initiated
2018-12-17
Root cause
Device Design
Terminated
2020-11-02
Distribution
US Nationwide distribution.
Recall number
Z-1145-2019
Recall status
Terminated
CDRH record ID
170148
FDA classified
2019-04-13
Recall event number
81859

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗