Open original FDA record ↗Recall record
- FDA identifier
- Z-1145-2020
- Company in FDA record
- Philips North America, LLC
- FDA classification date
- Feb 10, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction.
Source facts
- Action
- On 12/23/2019, the firm sent medical device correction letters to affected customers via certified mail. The firm will be contacting customers with affected systems. The firm will be implementing software updates in all affected systems to correct the issue which was made available by the end of December 2019. The user was asked to do the following: -Please suspend the use of Digital Subtraction Angiography images for the verification of the registration during the Live Guidance step. -Please ensure that all staff working with the VesselNavigator application are informed of the recall. If you need further information or support concerning this issue, please contact your local Philips representative or call Technical Support Line at 1-800-722-9377.
- Initiated
- 2019-12-23
- Root cause
- Software design
- Terminated
- 2022-11-17
- Distribution
- Worldwide Distribution. US nationwide, Canada, Argentina, Australia, Austria, Belgium, Brazil, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Latvia, Libya, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Philippines, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Uruguay.
- Recall number
- Z-1145-2020
- Recall status
- Terminated
- CDRH record ID
- 179087
- FDA classified
- 2020-02-10
- Recall event number
- 84708