This correction was for several software bugs which caused image format problems; incorrect ruler display; software crashes; failure to update pixel spacing; display of unattainable treatment table positions; failure to update gantry angles; wrong shifts for two of three axes of treatment table movement.
Source facts
Action
Software to fix the problems was written and software users were notified of the availability by e-mail on 04/07/2006. They could download the fixed software from the firm's internet website.
Initiated
2006-04-07
Root cause
Software design
Terminated
2011-04-19
Distribution
Customers in USA, Korea, Spain and UK.
Recall number
Z-1146-2007
Recall status
Terminated
CDRH record ID
53476
FDA classified
2007-08-08
Recall event number
38355