Open original FDA record ↗Recall record
- FDA identifier
- Z-1146-2010
- Company in FDA record
- Ventana Medical Systems Inc
- FDA classification date
- Mar 16, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect result: Light staining inconsistencies with lot number 680957 of the Ventana PATHWAY HER-2/neu (4B5) Rabbit Monoclonal Primary Antibody.
Source facts
- Action
- The recall was initiated on 09/17/2009 with Ventana Medical Systems forwarding Urgent Voluntary Medical Device Recall letters to all the affected customers advising them to discontinue the use of lot number 680957 and have their laboratory retest with product from a different lot of PATHWAY HER2 (4B5) any breast cases scored as 0, 1+, or 2+ that were not reflexed to an approved in situ hybridization (ISH) method. Customers were also instructed to contact Ventana Customer Support at 1-800-227-2155 and select 1 for Technical Support to speak to a Customer Support agent who will work with then to replace or credit any product from this specific lot number.
- Initiated
- 2009-09-17
- Root cause
- Nonconforming Material/Component
- Terminated
- 2010-04-22
- Distribution
- Worldwide distribution: USA, Austria, Denmark, Germany, Hungary, Netherlands, Sweden, United Kingdom, Brazil, Canada, India, Korea, Morocco & Thailand
- Recall number
- Z-1146-2010
- Recall status
- Terminated
- CDRH record ID
- 85961
- FDA classified
- 2010-03-16
- Recall event number
- 53597