Open original FDA record ↗Recall record
- FDA identifier
- Z-1146-2018
- Company in FDA record
- Vascular Insights, LLC
- FDA record creation date
- Mar 21, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain catheters could have a compromised catheter shaft due to a manufacturing defect which could result in some delamination of the catheter shaft including during introduction of the catheter into a vascular access introducer.
Source facts
- Action
- Vascular Insights sent an Urgent Voluntary Medical Device Recall letter to their customers. The letter identified the affected product, problem and the actions to be taken. Customers were instructed to 1. Do not use or distribute any recalled product. 2. Immediately remove all recalled product from your inventory. 3. Segregate recalled product in a secure location and follow the instruction herein to return the recalled product and receive replacement product. 4. Immediately forward a copy of this recall notification to any/all sites to which you may have distributed the affected product and provide directions to follow the instructions contained herein. Customers were asked to complete and return the enclosed Reply Verification Tracking Form.
- Initiated
- 2017-11-09
- Root cause
- Process control
- Terminated
- 2018-12-18
- Distribution
- U.S Nationwide in the states: AZ, CA, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, MO, NE, NJ, NY, OH, PA, SC, TN, TX, VA, WA
- Recall number
- Z-1146-2018
- Recall status
- Terminated
- CDRH record ID
- 161969
- FDA record created
- 2018-03-21
- Recall event number
- 79324