Z-1146-2021: ENROUTE Transcarotid Stent System REF SR-XXYY-CS

Recall record Z-1146-2021 for Silk Road Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1146-2021
Company in FDA record
Silk Road Medical Inc
FDA record creation date
Feb 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to complaints received that the tip/nose cone may detach from the Transcarotid Stent System.

Source facts

Action
On 01/13/2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification via email to customers informing them that product complaints have been received where the tip/nose cone detached from the stent delivery system during use and it is possible the resulting patient harm can range from minor intervention required to retrieve the detached tip/nose cone to embolization or stroke if the device failure goes undetected. Customers are instructed to: 1. Immediately removed the affected lots from hospital inventory and quarantine the products in a secure location until such time as the Recalling Firm can arrange for removal from their facility. The affected units will be permanently removed from commercial distribution. There is no need for action related to devices already used during a procedure. 2. Immediately upon receipt of the email notification, reply to the email to recall@silkroadmed.com and record the number of units for each affected lot 3. If not impacted products remain in their facility or a Recalling Firm's field representative has already removed the affected products, customer are to indicated this accordingly on the email reply. For any questions, contact the Recalling Firm at 855-410-8227 ext. 2110 Monday through Friday 8:00 a.m. to 5:00 p.m. Pacific Time or email recall@silkroadmed.com
Initiated
2021-01-13
Root cause
Under Investigation by firm
Terminated
2023-01-31
Distribution
U.S.: AZ, CA, CO, CT, DC, DE, FL, GA, HI, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: None
Recall number
Z-1146-2021
Recall status
Terminated
CDRH record ID
185523
FDA record created
2021-02-23
Recall event number
87196

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗