Open original FDA record ↗Recall record
- FDA identifier
- Z-1147-06
- Company in FDA record
- Alcon Laboratories, Inc.
- FDA classification date
- Jun 21, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Alcon has identified a software error associated with the use of the Measuring Mode (ruler tool) on the LADARVision4000 Excimer Laser Systems. Use of this feature after alignment of the horizontal reference line will negate compensation for cyclotorsion on CustomCornea procedures, possibly adversely affecting clinical outcomes.
Source facts
- Action
- A Medical Device Safety Alert dated August 1, 2005, has been issued to all consignees. Each consignee was issued a Response To Safety Alert form to acknowledge receipt of the Safety Alert. A Field Service Engineer will visit to install the revised software.
- Initiated
- 2005-08-01
- Root cause
- Other
- Terminated
- 2012-07-06
- Distribution
- There have been 285 units distributed in the US and 151 units distributed to International Alcon Affiliates.
- Recall number
- Z-1147-06
- Recall status
- Terminated
- CDRH record ID
- 40995
- FDA classified
- 2006-06-21
- Recall event number
- 32866