When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display values ten times greater than the correct values.
Source facts
Action
Software to fix the described problems was written and customers contacted by e-mail on 05/31/2005, that a software patch, titled 05312005 was available on the firm's internet website.
Initiated
2005-05-31
Root cause
Software design
Terminated
2011-04-19
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Recall number
Z-1147-2007
Recall status
Terminated
CDRH record ID
53480
FDA classified
2007-08-08
Recall event number
38358