Reason stated in the source
Ethicon discovered that specific lots of DERMABOND"PRINEO"System may not dry within the specified time after proper application, and thus may fall off.
Source facts
- Action
- Ethicon sent an URGENT MEDICAL DEVICE RECALL REMOVAL letter dated November 2017 to customers titled: "DERMABONDTM PRINEOTM SKIN CLOSURE SYSTEM (22 CM)". The letter identified the affected product, problem and action to be taken. The letter instructed customers to do not use or distribute any product which is subject to recall. Customers are asked to examine inventory, quarantine products that are subject to recall and contact the firm for return of product. Also, post the recall letter in a visible location at facility. If you have additional questions regarding this recall (removal) or to report any customer complaints, please contact Ethicon Customer Support Center at 1-877-ETHICON (1-877-384-4266). The Customer Support Center is open Monday through Friday, 7:30 AM to 6:30 PM ET.
- Initiated
- 2017-11-27
- Root cause
- Component change control
- Terminated
- 2019-04-30
- Distribution
- US Nationwide Distribution
- Recall number
- Z-1147-2018
- Recall status
- Terminated
- CDRH record ID
- 162112
- FDA record created
- 2018-03-21
- Recall event number
- 79355