Skip to content

Public FDA record

Z-1148-2007: biphasic LIFEPAK 500 automated external defibrillators (AEDs) with software version 4.4 or lower; Part No

Recall record Z-1148-2007 for Medtronic Emergency Response Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1148-2007
Company in FDA record
Medtronic Emergency Response Systems, Inc.
FDA classification date
Aug 4, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Reduced Shock- the device may deliver 100 Joule (J) to a patient, which is less than the recommended factory minimum default setting of 200 J.

Source facts

Action
On June 13, 2007 the recalling firm sent a letter dated June 11, 2007, titled URGENT - MEDICAL DEVICE CORRECTIVE ACTION. The letter advises of the situation, recommends to the consignee to set defibrillator to FIXED Sequence Protocol and to contact the firm for a FLEXIBLE Sequence Energy Protocol. The letter provides an instruction sheet and a response sheet to be faxed back to the company.
Initiated
2007-06-13
Root cause
Other
Terminated
2011-03-20
Distribution
Worldwide
Recall number
Z-1148-2007
Recall status
Terminated
CDRH record ID
53210
FDA classified
2007-08-04
Recall event number
38232

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗