Open original FDA record ↗Recall record
- FDA identifier
- Z-1148-2018
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Mar 20, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor.
Source facts
- Action
- Siemens sent an Urgent - Medical Device Correction letter dated January 5, 2018. Siemens will conduct an inspection of all affected Artis tables and verify the correct positioning of the bolt as well as the snap rings. The letter recommends the consignees follow Please follow the standard emergency procedures they have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. The Siemens service organization will contact the consignees to arrange a date to perform the corrective action. For further questions, please call (610) 448-6461.
- Initiated
- 2018-01-12
- Root cause
- Process control
- Terminated
- 2018-08-28
- Distribution
- US Distribution to the states of : AL, AR, CA, FL, IN, LA, MI, MO, NC, NY, PA, TX, and WA.
- Recall number
- Z-1148-2018
- Recall status
- Terminated
- CDRH record ID
- 162604
- FDA classified
- 2018-03-20
- Recall event number
- 79538