Open original FDA record ↗Recall record
- FDA identifier
- Z-1148-2020
- Company in FDA record
- Stryker Corporation
- FDA classification date
- Mar 2, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.
Source facts
- Action
- On July 23, 2019, a "Urgent Medical Device Recall Notification" was emailed to sales representatives. In addition to informing consignees on the cause of the recall, the following action was requested: 1. Check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your accounts. Response is required by August 15, 2019. 2.If affected product is found in your inventory, segregate the product and call Stryker customer service at 1-866-596-2022 or email smwest@stryker.com to arrange for product return and issuance of credit or replacement (upon availability). Also fill out Attachment A and return to endorecall@stryker.com. 3.If affected product is found at account(s), please complete both steps below: a.Provide attached memo (Attachment B) to the account(s) to acknowledge the removal of devices and send completed memo back to endorecall@stryker.com. b.Segregate the product and call Stryker customer service at 1-866-596-2022 or email smwest@stryker.com to arrange for product return and issuance of credit or replacement (upon availability). 4.If no product is found in your inventory or at your account(s), complete acknowledgement form on Attachment A and return it back to endorecall@stryker.com.
- Initiated
- 2019-07-19
- Root cause
- Nonconforming Material/Component
- Terminated
- 2024-02-02
- Distribution
- US: GA, NJ, NY, KY, MD, TX, UT, AZ, IL, ID, CO, OH, FL, AK, CA, NV, AR, ME, WI
- Recall number
- Z-1148-2020
- Recall status
- Terminated
- CDRH record ID
- 179068
- FDA classified
- 2020-03-02
- Recall event number
- 84664