Reason stated in the source
The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.
Source facts
- Action
- Firm emailed all consignees and instructed them to remove/discard affected wedges (bottles) and contact Medica for immediate replacement 1 Immediately discontinue use of the Wash1 wedges (REF 10680) (Lot# listed above) and discard any remaining kits/wedges. 2. Please discuss possible concerns about lipase results achieved using these lots of Wash wedges with your laboratory manager and/or clinician in order to decide on appropriate measures. Report any concerns to Medica for evaluation and review. 3.Submit claims for replacement of affected product to your authorized Medica EasyRA dealer. For further questions, please call (781) 541-7560.
- Initiated
- 2017-11-21
- Root cause
- Under Investigation by firm
- Terminated
- 2020-08-20
- Distribution
- Worldwide and US Distribution.
- Recall number
- Z-1149-2018
- Recall status
- Terminated
- CDRH record ID
- 161792
- FDA record created
- 2018-03-21
- Recall event number
- 79264