Open original FDA record ↗Recall record
- FDA identifier
- Z-1149-2021
- Company in FDA record
- Steris Corporation
- FDA record creation date
- Feb 24, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
Source facts
- Action
- In January, 2021, Steris will issue an Urgent Medical Device Correction notice to customers via letter for the OT1000 Series Orthopedic Surgical Table. Steris has identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
- Initiated
- 2021-01-21
- Root cause
- Process control
- Terminated
- 2021-11-10
- Distribution
- Domestic: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL IN, IW, KS, KY, LA, ME, MD, MI, MN, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, PR, SC, TX, UT, VT, VA, WA, WV, WI. Foreign: Canada, South Korea, Panama, Qatar, Thailand.
- Recall number
- Z-1149-2021
- Recall status
- Terminated
- CDRH record ID
- 185584
- FDA record created
- 2021-02-24
- Recall event number
- 87230