Z-1150-2013: EVOTECH Endoscope Cleaner and Reprocessor System, P/N 50004. Designed to automatically clean and high-le

Recall record Z-1150-2013 for Advanced Sterilization Products. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1150-2013
Company in FDA record
Advanced Sterilization Products
FDA classification date
Apr 18, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because they have received reports of smoke emanating from the system due to a circuit board malfunction that may occur in the Channel Control Manifold sub-assembly.

Source facts

Action
An Urgent Medical Device Field Correction letter dated 3/28/13 was sent to all customers who purchased the EVOTECH Endoscope Cleaner and Reprocessor (ECR). The letter informed the customers of the problems identified and the actions to be taken. Customers are informed that the failure can potentially cause the following outcomes for the cycles listed below: 1. In the Full Reprocessing Cycle (Cleaning and HLD): there is a potential for residual rinse water and/or alcohol to remain in the endoscope channels after the cycle completes. 2. In the HLD-Only Cycle: there is potential for an adequate amount of high-level disinfectant solution to be sent through some endoscope channels. If customers observe smoke, then they are instructed to: a. Power OFF the device. b. Report your incident to ASP Professional Services at (888) 783-7723 and select option #2. c. Do not use the device until it has been serviced by ASP. Customers are instructed to return attached buyer return card acknowledging their receipt of this communication. Customers are instructed to contact Stericycle at (866) 761-9566 for any questions related to the field correction.
Initiated
2013-03-28
Root cause
Under Investigation by firm
Terminated
2013-08-22
Distribution
Nationwide Distribution
Recall number
Z-1150-2013
Recall status
Terminated
CDRH record ID
117051
FDA classified
2013-04-18
Recall event number
64781

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗