Open original FDA record ↗Recall record
- FDA identifier
- Z-1150-2018
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Mar 20, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain systems may have leaking of the plastic coolant couplings, the liquid can enter into the flat panel detector leading to a failure of the detector. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
Source facts
- Action
- The firm, Siemens Healthineers, sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated 1/5/2018 to its consignees. The letter describes the product, problem and actions to be taken. The consignees were instructed to follow their standard emergency procedures in place in the event of system failures; to consider reinforcing these emergency procedures until the update has been completed. Siemens will replace all affected plastic couplings with metal couplings. The Siemens service organization will contact the consignees to arrange a date to perform this corrective action.If the device has been sold and is therefore no longer in your possession, please forward this field correction notice to the new owner. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
- Initiated
- 2018-01-05
- Root cause
- Process control
- Terminated
- 2018-08-23
- Distribution
- US Distribution in the following U.S. states: AZ, CA, FL, IA, IL, LA, MI, NC, NJ, OH, OK, and TX.
- Recall number
- Z-1150-2018
- Recall status
- Terminated
- CDRH record ID
- 162606
- FDA classified
- 2018-03-20
- Recall event number
- 79540