Open original FDA record ↗Recall record
- FDA identifier
- Z-1151-2017
- Company in FDA record
- Vertebral Technologies, Inc.
- FDA record creation date
- Feb 8, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Fit issues have been identified with the Tail Traction Tool (PN 9116) when used with the Interfuse T tailed implant. The InterFuse T Tail may not fit into the Thru Hole in the instrument.
Source facts
- Action
- Consignees were sent via email a VTI "Advisory Notice" letter. The letter listed the Concern, Impact, Alternate Methods, Affected Lots, and Your Responsibility. Consignees were requested to acknowledge receipt of this notice, Discontinue use of the Tail Traction Tool when implanting the InterFuse T tailed implant, Notify individuals that may potentially use the device and alternate module assembly methods. Advise consignees to contact Customer Service (1-877-912-5401 in US , +-952-912-5400 outside US or customerservice@vti-spine.com to obtain a replacement instrument.
- Initiated
- 2017-01-13
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-10-05
- Distribution
- US: IL, WI, OH, NY, MT, FL, MD, IN, TN, TX, MN. OUS: Italy, Taiwan Spain, Belgium, Canada, Netherlands, Sweden, Switzerland, Portugal, Brazil. Germany, Ireland,
- Recall number
- Z-1151-2017
- Recall status
- Terminated
- CDRH record ID
- 152615
- FDA record created
- 2017-02-08
- Recall event number
- 76275