When ISOLOC version 6.0 and 4.2 distance unit is set to centimeter, the localization coordinates were rounded to the nearest millimeter which could affect positioning by 0.5 mm.
Source facts
Action
Software to fix the described problems was written and customers contacted by e-mail on 04/28/2005, that a software patch, titled 04282005, was available on the firm's internet website.
Initiated
2005-04-28
Root cause
Software design
Terminated
2011-04-20
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Recall number
Z-1152-2007
Recall status
Terminated
CDRH record ID
53482
FDA classified
2007-08-08
Recall event number
38360