When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm.
Source facts
Action
Software to fix the described problem was written and customers contacted by e-mail on 04/11/2005, that a software patch, titled 04112005, was available to their customers on the firm's internet website.
Initiated
2005-04-11
Root cause
Software design
Terminated
2011-04-19
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Recall number
Z-1153-2007
Recall status
Terminated
CDRH record ID
53485
FDA classified
2007-08-08
Recall event number
38361