When ISOLOC version 6.0 software was used in the Anatomic Landmark mode, the couch moves display in the Couch Translator window was incorrect.
Source facts
Action
Software to fix the described problem was written and customers contacted by e-mail on 12/30/2004 that a software patch, titled 12302004 was available to their customers on the firm's internet website.
Initiated
2004-12-30
Root cause
Software design
Terminated
2011-04-19
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Recall number
Z-1154-2007
Recall status
Terminated
CDRH record ID
53488
FDA classified
2007-08-08
Recall event number
38363