Skip to content

Public FDA record

Z-1154-2009: Alm Prismatic Surgical Light 5004 SF; Maquet SA. ALM PRISMATIC (PRC) Surgical Lights are used to provide

Recall record Z-1154-2009 for MAQUET Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1154-2009
Company in FDA record
MAQUET Inc.
FDA classification date
Apr 12, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The screws holding the Prismatic triple surgical lights may break due to dynamic multidirectional bending stress. The firm has received one complaint report.

Source facts

Action
An "Urgent Device Correction" letter dated February 23, 2009 was mailed by FedEx Tracking to all affected consignees. The letter described the issue, potential hazard and recommended precaution. Customers were instructed to complete the enclosed "Customer Response Fax Form" and return it via fax (1-732-667-1903) to Maquet, Inc. Maquet, Inc. will schedule repair service of the affected units once the "Customer Response Fax Form" is received. Direct questions to your local Maquet, Inc. Representative.
Initiated
2009-02-23
Root cause
Device Design
Terminated
2009-09-29
Distribution
Nationwide Distribution.
Recall number
Z-1154-2009
Recall status
Terminated
CDRH record ID
79673
FDA classified
2009-04-12
Recall event number
51297

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗