Reason stated in the source
Inaccurate guidelines: The BE9C is an optional, small radius Bi-planar convex probe intended for use with the LOGIQ 3 Expert system in Urology application. An attempt by the user to activate the biopsy guidelines while this probe is in use will cause incorrect guidelines to be displayed on the image. The incorrect guidelines are recognized by their orientation that is left/right reversed from t
Source facts
- Action
- GE sent consignees a "GE Urgent Medical Device Correction" letter dated 8/14/07. The letter informed the customers about the problem, offer a short-term workaround and informed them that a representative will schedule a visit to upgrade the software.
- Initiated
- 2007-09-04
- Root cause
- Software change control
- Terminated
- 2012-05-29
- Distribution
- Worldwide Distribution - USA (NY & AL), Korea, China, India, Brazil, Mexico, Germany, Greece, Italy, Poland, Russia, Saudi Arabia, Spain, and Czech Republic.
- Recall number
- Z-1155-2008
- Recall status
- Terminated
- CDRH record ID
- 64472
- FDA classified
- 2008-06-12
- Recall event number
- 44884