Z-1155-2018: J-Plasma(R) Handpiece, Catalog Numbers: GS-018C, GS-270C, BVX-270B

Recall record Z-1155-2018 for Bovie Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1155-2018
Company in FDA record
Bovie Medical Corporation
FDA record creation date
Mar 21, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists for patient fluid backflow into the handpiece and the generator receptacle leading to risk of cross-contamination if the generator is reused.

Source facts

Action
The firm initiated their recall on 02/08/2108 by email to distributors, and letter to the remaining consignees. Distributors were directed as follows: "We ask that you immediately: 1. Quarantine your inventory of the J-Plasma Handpieces o Provide us with a list of catalog numbers and quatities that you have on your possession using the attached Inventory List template. o Email the information to me at Rubiela.Maldonado@boviemed.com Note: as soon as we receive the information from you we will ship the box labels and insert stuffer notices for the stock that you have on your possession. o After receiving these labels, please follow the attached instructions on how to add each of the insert stuffer notice and the outer box label to each of the boxes that you have on your possession. o Document the number of labels/stuffers added to each of the products on the Inventory List provided, sign and date to confirm this was performed. o Return the signed document to me via email at Rubiela.maldonado@boviemed.com. 2. Inform your customers about the correction: o Identify and contact your customers that have received the product affected by this correction. o Provide them with a copy of the Urgent Medical Device Correction Template attached to this letter. o Track and document the customer responses to ensure 100% effectiveness. o Provide us a summary of the effectiveness of this correction (number of customers impacted and number of responses received)." The remaining consignees (users of the device) were instructed as follows: "ACTIONS: At the end of each procedure, inspect the handpiece cable plug and generator receptacle for the presence of fluids. If fluids are present, immediately discontinue use of the generator. In addition to following the J-Plasma handpiece Instructions for Use, immediately begin to follow the supplemental instructions provided below to ensure that the J-Plasma handpiece cable plug is properly and co
Initiated
2018-02-08
Root cause
Nonconforming Material/Component
Terminated
2020-09-17
Distribution
worldwide
Recall number
Z-1155-2018
Recall status
Terminated
CDRH record ID
162211
FDA record created
2018-03-21
Recall event number
79398

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗