Installation problem: there is a risk that the monitor may not be adequately secured to the stand, resulting in the monitor falling off which may cause injury.
Source facts
Action
St. Jude Medical received a recall notice dated January 16, 2009 from its supplier, Smiths Medical PM, Inc. St. Jude Medical prepared and sent their own recall letter dated February 13, 2009 to affected customers. The recall letter described the issue and possible associated risks and provided instructions to customers. A St. Jude Medical representative will contact customers to facilitate replacement of the fasteners. Questions should be directed to your St. Jude Medical sales representative.
Initiated
2009-02-13
Root cause
Nonconforming Material/Component
Terminated
2009-05-04
Distribution
Nationwide Distribution -- CA, NV, and FL.
Recall number
Z-1156-2009
Recall status
Terminated
CDRH record ID
79130
FDA classified
2009-04-10
Recall event number
51195