Product failure-- Product may cease operation as a result of potential speed control board failure when the device is shut off, and may not turn back on. A component on the circuit board fails unexpectedly because of a power surge, causing a diode to burn out.
Source facts
Action
The firm has issued notification letters via Federal express with reply verification tracking forms. A follow up letter and telephone survey as needed will be sent to follow up with non-responders. The firm is asking its consignees to return any recalled product.
If there are any questions, please call 1.408.935.4952.
Initiated
2009-09-22
Root cause
Nonconforming Material/Component
Terminated
2010-12-27
Distribution
US and to Spain, France, Germany, Italy, Russia, Poland, the Netherlands, Israel and Belgium
Recall number
Z-1156-2010
Recall status
Terminated
CDRH record ID
85564
FDA classified
2010-03-17
Recall event number
53463