Skip to content

Public FDA record

Z-1156-2021: HAMILTON HEATER SHAKER, 115 VAC/ 230 VAC, 750 VA, Power Consumption 41V/ 100W (max.), Relative Humidity 1

Recall record Z-1156-2021 for Hamilton Co. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1156-2021
Company in FDA record
Hamilton Co
FDA record creation date
Feb 27, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a risk that the device can start a fire if flammable liquids contact and enter the device.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION notifications was sent to the consignees via email of 01/15/2021. Hamilton Heater Shaker upgrades were started on 15-JAN- 2021. A device upgrade has been developed to mitigate the risk of the device catching fire. The upgrade consists of a plastic tray within the body of the Hamilton Heater Shaker and a redesigned base plate. These upgrades prevent liquid from reaching the electrical components of the device. The recall notification states: " Hamilton Company Field Service will contact your organization to schedule a time to upgrade your Hamilton Heater Shaker(s) to mitigate this risk. For customers who perform their own field service, a zero-dollar purchase order can be submitted to RoboticsOEMsales@hamiltoncompany .com to obtain upgrade kits. The upgrade consists of a plastic tray within the body of the Hamilton Heater Shaker and a redesigned base plate. These upgrades prevent liquid from reaching the electrical components of the device. This upgrade will be supplied free-of-charge and does not impact the functionality of the device. Until the upgrade is performed, it is crucial to observe the following safety best practices: o Methods should include labware touch-off presence checks to ensure the labware is ready to receive liquid. o On-deck waste and fluid collection devices should be placed at least one track away from Hamilton Heater Shakers or any other heating device. Recipients of this notice shall forward it to all customers to whom they have sold Hamilton Heater Shakers. Please complete and return the attached Acknowledgement and Receipt Form to productrecalls@hamiltoncompany.com . On this form, please indicate if you have experienced a fire on your device. If you experience a fire on your device in the future, please notify your Hamilton representative immediately and cease using the device. Please contact productrecalls@hamiltoncompany.com or call 800-527-5269 between 7:00am and 3:30pm P
Initiated
2021-01-15
Root cause
Device Design
Distribution
US - AL, AR, AZ, CA, CO, CT, DC. DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA. MD, ME. MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI AND WY OUS - AUSTRALIA, AUSTRIA, BRAZIL , CANADA , CHILE, CHINA, , COLOMBIA , FINLAND , GERMANY , ITALY , JAPAN , JPN , KOREA, SOUTH, PARAGUAY , PERU , PHL , ROMANIA , ROU , SINGAPORE, SWITZERLAND , SWITZERLAND , THAILAND, TWN , UNITED KINGDOM, AND URUGUAY
Recall number
Z-1156-2021
Recall status
Open, Classified
CDRH record ID
185562
FDA record created
2021-02-27
Recall event number
87219

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗