Skip to content

Public FDA record

Z-1157-2007: The ClearGlide EVH small with Scissors. The kit includes three Precision Bipolar Devices, an Optical Vess

Recall record Z-1157-2007 for Datascope Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1157-2007
Company in FDA record
Datascope Corporation
FDA classification date
Aug 9, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Datascope has experienced a higher than usual number of reports involving the Precision Bipolar Device. Reports include the locking or sticking of the buttons of the device which operate the jaws, particularly when a side load is placed on the shaft of the device.

Source facts

Action
Letters were sent to all hospitals, (contact: Hospital Administrators) the week of May 7, 2007 (certified mail, Return Receipt requested) with Urgent Medical Device Field Correction. The letter requests that all potential users of the devices at the institution be mdae aware of the situation. The letter instructs clinicians what to do if they experience problems with locking buttons of the Precision Device and also if the device is clamped closed on a vein branch and cannot be opened. Institutions are also requested to sign and return a postal card acknowledging receipt of the Field Correct notification and Datascope provides telephone numbers for customer service.
Initiated
2007-05-07
Root cause
Other
Terminated
2008-03-04
Distribution
Worldwide, including USA, Belgium, the Netherlands, and Germany.
Recall number
Z-1157-2007
Recall status
Terminated
CDRH record ID
52710
FDA classified
2007-08-09
Recall event number
38005

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗