Skip to content

Public FDA record

Z-1157-2020: Implant Direct, CE 0086, REF: 855710, Legacy 3 Impant, Surface: SBM, Qty:1, Size:5.7mmD x 10mmL, 5.7mmD P

Recall record Z-1157-2020 for Implant Direct Sybron Manufacturing LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1157-2020
Company in FDA record
Implant Direct Sybron Manufacturing LLC
FDA classification date
Feb 5, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Through a customer complaint and and internal investigation it was determined that some implant packages contain the incorrect cover screw and extender. This could result in delay in surgery and/or a potential new surgical procedures to place the correct parts. The level of severity varies based on a one or two stage surgery.

Source facts

Action
On October 11, 2019 the firm sent an "URGENT: MEDICAL DEVICE RECALL" letter via FedEx overnight priority mail. The letter advised the consignees of the recall. In addition, the letter states: We request that you take the following actions immediately:. 1. Please review your inventory for the affected product. If located, please quarantine the affected product and return the product to us according to the instructions on the attached form. 2. Please complete and return the Acknowledgement Form within fortyeight (48) hours of receipt of this notification, whether you have affected product or not. 3. If you are an authorized Implant Direct Sybron Manufacturing LLC distributor, we request that you identify those customers that may have been shipped the affected product lot and contact these customers to inform them of this issue within fortyeight (48) hours of receipt of this notification. 4. If you have any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425. Customer Care is available Monday through Friday, between the hours of 5am-5pm PST. If you or your patients have experienced any issues as a result of the affected products noted in this URGENT: MEDICAL DEVICE RECALL HHE 855710 100419 Rev 01 Status: RELEASED printed 10/10/2019 2:27:23 PM by Jose Trejo www.implantdirect.com | 888-649-6425 communication, you may voluntarily report the incident to the FDA through the MEDWATCH reporting system at the following: http://www.fda.gov/Safety/MedWatch/HowToReport/ucm053074.htm
Initiated
2019-10-11
Root cause
Mixed-up of materials/components
Terminated
2024-04-12
Distribution
US: AR, AZ, CA, ID, IL, IN, KY, MD, ND, NH, NJ, NV, NY, Oh, OR, PA, TX, UT, WA, WI, WV, GA,MO
Recall number
Z-1157-2020
Recall status
Terminated
CDRH record ID
179406
FDA classified
2020-02-05
Recall event number
84777

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗