Open original FDA record ↗Recall record
- FDA identifier
- Z-1158-2020
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA record creation date
- Feb 12, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Alarm flags that have a "<" symbol in front of them are not displayed on the cobas infinity Validation screen. This issue affects only customers that are doing manual validation on the cobas infinity Validation screen.
Source facts
- Action
- On 1/7/2020, the firm distributed Urgent Medical Device Correction letters to affected consignees. Customers were asked to do the following: " Be aware of the incorrect display of the alarm flags on the cobas infinity Validation screen if an alarm has a < symbol in front of it and if doing manual validations. " It is not recommended to use the alarm mapping function on the cobas infinity until your system is patched. If your laboratory does use it, please call the Roche Support Network Customer Support Center at 1-800-526-2272 for further information. " Complete the customer response form and return it to the recalling firm. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272 if you have questions about the information contained in this UMDC.
- Initiated
- 2020-01-07
- Root cause
- Under Investigation by firm
- Terminated
- 2020-10-15
- Distribution
- Distributed to customers in AK, AL, CA, GA, HI, IA, IL, KY, MA, MI, MN, MO, MS, MT, NJ, NY, OR, SC, and TN.
- Recall number
- Z-1158-2020
- Recall status
- Terminated
- CDRH record ID
- 179073
- FDA record created
- 2020-02-12
- Recall event number
- 84697