- FDA identifier
- Z-1159-06
- Company in FDA record
- Medtronic Gastroenterology / Urology
- FDA classification date
- Jun 22, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Mislabeling- The 8 ports of the manometric catheter were incorrectly labeled 1,3,5,7,8,2,4,6 from the distal to the proximal end of the catheter. The correct port configuration is sequentially 1 through 8, from the distal to the proximal end of the catheter.
Source facts
- Action
- Recall initiated to International accounts on 04/05/06 and within US on 04/06/06 via an Urgent Recall Notification. The letter informs them of the manufacturing defect, requests return for credit of any unused inventory and requests accounting of all previously used catheters. Distributors are instructed to sent a letter to their customers to request product return.
- Initiated
- 2006-04-06
- Root cause
- Other
- Terminated
- 2007-01-16
- Distribution
- Worldwide Distribution-including states of IL, NC, NY. and countries of Austria, Chile, Finland, Germany, Greece, Hong Kong, Italy, Japan, Lebanon, Norway, Peru, Portugal, Saudi Arabi, Spain
- Recall number
- Z-1159-06
- Recall status
- Terminated
- CDRH record ID
- 45292
- FDA classified
- 2006-06-22
- Recall event number
- 35094