Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images may display dark and not suitable for diagnostic. Due to this software error customers have to repeat x-ray exposure to receive diagnostic images.
Source facts
Action
Siemens sent Update Instruction XP057/11/S by mail on February 15, 2012, which releases software update package VC10D to remedy the described issues.
For questions regarding this recall call 610-219-4834.
Initiated
2012-02-15
Root cause
Software design
Terminated
2014-01-14
Distribution
Nationwide Distribution including AK, CA, CO, FL, GA, IL, IN, IA, KY, MD, MA, MO, NE, NY, NC, OH, OR, PA, SC, TX, and UT.
Recall number
Z-1159-2012
Recall status
Terminated
CDRH record ID
107662
FDA classified
2012-03-05
Recall event number
61251