Skip to content

Public FDA record

Z-1160-2021: Digitial Medica X-ray Imagining System, Model: uDR 596i, RX, CE0123, UDI:(01) 06971576833023 - Product Us

Recall record Z-1160-2021 for Shanghai United Imaging Healthcare Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1160-2021
Company in FDA record
Shanghai United Imaging Healthcare Co., Ltd.
FDA record creation date
Mar 2, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the X-Ray tube head (Touch screen interface) to perform an uncommanded motion. This could result in a collision with the patient.

Source facts

Action
On 01/25/2021, Shanghai United Imaging sent an Urgent: Medical Device Correction via: Email. In addition to informing consignees about the recall, the firm asked customers to take the following actions: 1. Once the RHA motor driver is powered on, and while the RHA motor is looking for the zero-position (before its first electric movements), please do not apply any external force to the TUI. 2. For further questions, please contact your United Imaging Service Engineer or call the United Imaging Customer Contact Center at our 24-hour hotline (855) 221-1552 (toll free) or (832) 699-2799 (direct). 3. Adverse reactions or quality problems experienced with the use of this product may be reported to FDAs MedWatch Adverse Event Reporting program either online, by regulatory mail or by fax. 4. Notify all users of the affected products via this Customer Notification Letter. 5. Please completed the Receipt of Customer Notification Letter form and return to United Imaging.
Initiated
2021-01-25
Root cause
Device Design
Terminated
2022-05-03
Distribution
US Nationwide distribution in the state of TX.
Recall number
Z-1160-2021
Recall status
Terminated
CDRH record ID
185730
FDA record created
2021-03-02
Recall event number
87289

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗