There is a potential for the X-Ray tube head (Touch screen interface) to perform an uncommanded motion. This could result in a collision with the patient.
Source facts
Action
On 01/25/2021, Shanghai United Imaging sent an Urgent: Medical Device Correction via: Email. In addition to informing consignees about the recall, the firm asked customers to take the following actions:
1. Once the RHA motor driver is powered on, and while the RHA motor is looking for the zero-position (before its first electric movements), please do not apply any external force to the TUI.
2. For further questions, please contact your United Imaging Service Engineer or call the United Imaging Customer Contact Center at our 24-hour hotline (855) 221-1552 (toll free) or (832) 699-2799 (direct).
3. Adverse reactions or quality problems experienced with the use of this product may be reported to FDAs MedWatch Adverse Event Reporting program either online, by regulatory mail or by fax.
4. Notify all users of the affected products via this Customer Notification Letter.
5. Please completed the Receipt of Customer Notification Letter form and return to United Imaging.
Initiated
2021-01-25
Root cause
Device Design
Terminated
2022-05-03
Distribution
US Nationwide distribution in the state of TX.
Recall number
Z-1160-2021
Recall status
Terminated
CDRH record ID
185730
FDA record created
2021-03-02
Recall event number
87289