Open original FDA record ↗Recall record
- FDA identifier
- Z-1161-06
- Company in FDA record
- bioMerieux Inc
- FDA classification date
- Jun 22, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software problem with scanner inter-character delay which may result in false positive or false negative results after incorrect scanning of bottle ID's
Source facts
- Action
- The recalling firm telephoned customers who ordered the MB bottles and use the OBSERVA software ver. 2.0 to inform them of the issues. Recall letters were issued to all customers via overnight mail on 5/22/06 informing them a field representative would be visiting to make a short-term correction to the system. The letter also informed the customer a permanent correction to the software was scheduled for release later in the year.
- Initiated
- 2006-03-28
- Root cause
- Other
- Terminated
- 2008-03-07
- Distribution
- WorldwideDistribution--Nationwide to medical centers. Government distribution was made to MN and CA, and military distribution was made to OK. Foreign distribution was made to Thailand, The Netherlands, Germany, Greece, Portugal, and France.
- Recall number
- Z-1161-06
- Recall status
- Terminated
- CDRH record ID
- 46301
- FDA classified
- 2006-06-22
- Recall event number
- 35518