Skip to content

Public FDA record

Z-1161-06: OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to B

Recall record Z-1161-06 for bioMerieux Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1161-06
Company in FDA record
bioMerieux Inc
FDA classification date
Jun 22, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software problem with scanner inter-character delay which may result in false positive or false negative results after incorrect scanning of bottle ID's

Source facts

Action
The recalling firm telephoned customers who ordered the MB bottles and use the OBSERVA software ver. 2.0 to inform them of the issues. Recall letters were issued to all customers via overnight mail on 5/22/06 informing them a field representative would be visiting to make a short-term correction to the system. The letter also informed the customer a permanent correction to the software was scheduled for release later in the year.
Initiated
2006-03-28
Root cause
Other
Terminated
2008-03-07
Distribution
WorldwideDistribution--Nationwide to medical centers. Government distribution was made to MN and CA, and military distribution was made to OK. Foreign distribution was made to Thailand, The Netherlands, Germany, Greece, Portugal, and France.
Recall number
Z-1161-06
Recall status
Terminated
CDRH record ID
46301
FDA classified
2006-06-22
Recall event number
35518

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗