Z-1161-2021: ACIST Kodama Intravascular Ultrasound Catheter The Kodama Intravascular Ultrasound Catheter is a compon

Recall record Z-1161-2021 for Acist Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1161-2021
Company in FDA record
Acist Medical Systems
FDA classification date
Mar 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Test results from the manufacturing line found a piece of damaged o-ring in an unexpected section of the catheter. Further testing indicated that pieces (>200 micron) of damaged o-ring had the potential to be flushed out of the catheter. ACIST is confirming the source of the failure mode to assure the quality and reliability of the Kodama catheter. There have been no related field reports related to this incident, nor any evidence or report of patient injury or adverse health consequence.

Source facts

Action
The firm, ACIST, sent an, "URGENT: MEDICAL DEVICE RECALL" letter and response form dated Jan 22, 2021 to customers on Jan. 22, 2021. The letter describe the product, problem and actions to be taken. The customers were instructed to do the following: to complete all of the steps outlined below and return the completed Recall Response Form to Stericycle by e-mail: acistmedical8961@stericycle.com or fax to 877-576-9366. 1. Check your inventory of Kodama HD-IVUS Catheter 2. Record quantities of each lot in the Response Form 3. Remove the affected lots from your inventory. 4. Use the enclosed, prepaid return label to return your affected product including a copy of the response form with the product. If you need additional labels, please contact Stericycle at 877-576-9382. If you have received any reports of illness, injury or other health consequence related to the use of product please contact Customer Support: Customer.Support@acistmedical.com Please forward this notice to those who need to be aware within your organization. If you have any further questions or concerns, please contact Stericycle at 877-576-9382.
Initiated
2021-01-22
Root cause
Device Design
Terminated
2023-10-30
Distribution
Worldwide - US Nationwide Distribution in the states of AL, AR, CA, CO, FL, IL, KS, KY, LA, MD, MI, MO, NC, NJ, NY, OK, PA, RI, TN; In the countries of India, Italy, Japan, Poland, and United Arab Emirates.
Recall number
Z-1161-2021
Recall status
Terminated
CDRH record ID
185616
FDA classified
2021-03-04
Recall event number
87254

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗