Z-1162-06: Leica M520 Surgical Microscope; a surgical microscope with a 6:1 motorized zoom magnification, 207-470 mm

Recall record Z-1162-06 for Leica Microsystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1162-06
Company in FDA record
Leica Microsystems, Inc.
FDA classification date
Jun 24, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Patient Burns-The use of the surgical microscope at high light intensity and short working distances can result in patient burns.

Source facts

Action
Leica Microsystems faxed a memo entitled 'M520 Heat Issues' to all IMRs on 11/15/05, informing them that the potential to burn exposed patient skin at the closer working distances with the high intensity of light exists with the M520 optics carrier and/or new optichrome light source. The IMRs were sent retrofit packets which consisted of a current User Manual with 'Caution: The light source in this unit may be too bright at shorter focal lengths for surgeon and patient. Please start with light low and turn up till the surgeon has an excellent optical image.'; a sticker reading 'Warning: Maximum Light at Short Distance Can Injure Patient See User Manual', to be placed on the stand near the light intensity controller; a laminated card entitled 'Setting the Illumination Intensity' to be hung on the stand with the enclosed wire tie for in-service training; and a training verification form. The IMRs were instructed to visit each of the end user locations on the enclosed list, installing the stickers, providing the new User Manual and providing the staff that uses the microscope another in-service, stressing the proper use of the illumination system. Upon completion of the in-service, the IMRs will complete the training verification form and have a staff member from the OR sign the form verifying that the warning labels were applied and that the staff was trained, and that the user manual was received. The IMRs will then return the verification form to Leica Microsystems.
Initiated
2005-04-12
Root cause
Other
Terminated
2008-05-09
Distribution
Nationwide
Recall number
Z-1162-06
Recall status
Terminated
CDRH record ID
45967
FDA classified
2006-06-24
Recall event number
35385

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗