Skip to content

Public FDA record

Z-1163-2021: TECNIS Toric 1-Piece IOL, Model Number: ZCT150 +23.5D SE 1.50D CYL - Product Usage: intended to be placed

Recall record Z-1163-2021 for Johnson & Johnson Surgical Vision Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1163-2021
Company in FDA record
Johnson & Johnson Surgical Vision Inc
FDA record creation date
Mar 3, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the release of nonconforming Intraocular Lenses (IOLs).

Source facts

Action
On 01/27/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via confirmed mail to customers informing them that due to investigation results, following a customer complaint, which confirmed a hyperopic patient outcome post implantation for an Intraocular Lens (IOL). The investigation for this complaint determined that this occurred due to inadvertent release of nonconforming product (Product labeled as 23.5D but measured to be 17.5D). Customers are instructed to: 1) Compare their inventory against the attached list on page 3. The IOL serial number is displayed on the end of each individual unit carton (see page 3 for label example). 2) Discontinue using and remove from their inventory all affected IOLs. 3) Complete and return the attached Customer Reply Form to acknowledge receipt of this Recall Notice, even if they have no inventory. The Recalling Firm requires this information for reconciliation purposes with regulatory agencies. The Recalling Firm has pre-filled the Customer Reply Form with the IOL serial number(s) that was(were) shipped to them and marked accordingly where product has already been returned. The completed Customer Reply Form should be emailed to RegCompliOne@its.jnj.com or faxed to 1-714-460-0466 within 3 business days of receipt of the recall notification letter. 4) Should customers have product to return with the serial number listed on page 3, they are asked to contact Customer Support at 1-877-266-4543 to arrange pick up of affected product. 5) The Recall Notification Letter should be shared with anyone who needs to be aware within the customer's organization or to any organization where the potentially affected products have been transferred. Any questions contact the Public Contact/Director, Quality Assurance at 949-480-7526 or email jturpins@its.jnj.com
Initiated
2021-01-27
Root cause
Process change control
Terminated
2022-11-09
Distribution
U.S. Nationwide distribution in the states of AL, AZ, CA, FL, GA, IL, LA, MI, MN, NC, NY, OH, PA, SC, TN, TX, UT and WA.
Recall number
Z-1163-2021
Recall status
Terminated
CDRH record ID
185807
FDA record created
2021-03-03
Recall event number
87320

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗