Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large volumes of data are being reconstructed on the secondary console.
Source facts
Action
Customer notification was sent and a software upgrade will be issued to all customers affected by the field correction.
If you have questions, please call (800) 521-1968.
Initiated
2009-01-08
Root cause
Component design/selection
Terminated
2012-06-01
Distribution
Nationwide
Recall number
Z-1164-2010
Recall status
Terminated
CDRH record ID
84082
FDA classified
2010-03-19
Recall event number
52825
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.