Open original FDA record ↗Recall record
- FDA identifier
- Z-1164-2013
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Apr 23, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
GE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging (MPI) studies may create artifacts in attenuation corrected images. This may lead to a potential for misdiagnosis.
Source facts
- Action
- GE Healthcare sent an Urgent Medical Device Correction letter dated October 5, 2011, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Consignees were instructed to contact their local GE Healthcare Service Representative if they have questions or concerns. They may also contact GE Healthcare Call Center , United States or Canada at 800-437-1171. Other countries 262-896-2890. For questions regarding this recall call 262-513-4122.
- Initiated
- 2011-10-06
- Root cause
- Software design
- Terminated
- 2014-02-04
- Distribution
- Worldwide Distribution - USA including GA, and NY; and countries of: Canada, Israel, Italy and Switzerland.
- Recall number
- Z-1164-2013
- Recall status
- Terminated
- CDRH record ID
- 117103
- FDA classified
- 2013-04-23
- Recall event number
- 64804