Open original FDA record ↗Recall record
- FDA identifier
- Z-1164-2021
- Company in FDA record
- G & H Wire Company
- FDA record creation date
- Mar 3, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The text on the patient pack is correct and the product itself remains unchanged. FlexMedics does not expect device failures; Instead, there is an opportunity for misuse of the device if the label is not read. If the band is too small, the treatment may be more aggressive. It is important to note that elastic band force decreases over time. If the band is too small and still applied, the band could break. Both circumstances could result in temporary patient discomfort.
Source facts
- Action
- On 1/21/2021, G&H Orthodontics issued a Product Recall Notice to customers via email notifying affected users of a voluntary recall for FlexMedics Patient Elastic Packs. A labelling change that resulted in the color scheme of the FlexMedics Patient Packs product to be updated which was not intended and could result in consumer confusion and/or patient discomfort.
- Initiated
- 2021-01-21
- Root cause
- Process control
- Terminated
- 2021-09-20
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, CT, FL, IL, KY, MI, MN, MS, MT, NH, NJ, NY, NC, TN, TX, UT, VT, WA, WV, AR, IL, and the countries of Spain, Netherlands, New Zealand, Germany, France, Czech Republic, Greece, Poland, Thailand, Belgium, Denmark, United Kingdom, Canada.
- Recall number
- Z-1164-2021
- Recall status
- Terminated
- CDRH record ID
- 185621
- FDA record created
- 2021-03-03
- Recall event number
- 87248