Open original FDA record ↗Recall record
- FDA identifier
- Z-1164-2025
- Company in FDA record
- Microbiologics Inc
- FDA classification date
- Feb 12, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.
Source facts
- Action
- Microbiologics issued and Urgent Medical Device Recall notice to its consignees on 01/20/2025 via email. The notice explained the issue and requested the following actions be taken: INSTRUCTIONS: 1. REVIEW lab procedures to understand how this information affects your usage 2. DISCARD affected product. 3. COMPLETE the response form provided. 4. RETURN the response form to recall@microbiologics.com 5. KEEP this letter for your records. 6. NOTIFY other facilities or customers to whom the product may have been distributed or transferred. 7. CONTACT Microbiologics if a replacement kit is needed. Please contact our Recall Support team at 320.229.7080 or recall@microbiologics.com with questions or concerns.
- Initiated
- 2025-01-20
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.
- Recall number
- Z-1164-2025
- Recall status
- Open, Classified
- CDRH record ID
- 212183
- FDA classified
- 2025-02-12
- Recall event number
- 96166